Product introduction:
MOBILE X-RAY MACHINE MODEL SF70A
MEDICAL DIAGNOSTIC X-RAY EQUIPMENT
70mA x-ray equipment Main purpose and features:
This X-ray apparatus features fullwave rectification and modular
X-ray tube head. The frae of the unit, of cantilever construction, is
Compact in design easy of movement. The machine can be used for
Radiography in hospital wards or operating rooms.
Main Technical Data :
Maximum ratings
kV S mA |
0.06-1.0 | 1.2-2.5 | 3.2-6.3 |
30 | 90 | 90 | 90 |
50 | 90 | 80 | / |
70 | 70 | / | / |
1.Power requirements
Voltage:180-240V
Frequency:50Hz
Internal resistance: ≤1Ω
Rating:5k VA
Current:Maimum15A for radiography
2.Time:0.08 to 6.32 steps electronic
3.Maximum height from X-ray tube focus to floor>1820mm
4.Minimum height from X-ray tube focus to floor<550mm
5.Collimator:Maximum film size at 1000mm Focal distance: 430×430mm
6.Maximumermote control distance:6m
7.X-ray tube focus:2.6×2.6
8.Unit moving force:< 5kg
9.Net weight of the unit:90kg
400-686-9889
ADDRESS:565 jujiaqiao Road, Pudong New Area, Shanghai, China
EMAIL:2277936265@qq.com
TEL:021-62500690